Veelgestelde vragen

Foutmeldingen worden in Engels weergeven:

The control Unit had detected a problem calibrating the pneumatic compresssion circuit on start-up.

Solution(s) error 01

  • This is most likely to accur if you have restarted the System with an inflated Wrap attached.
  • Detach the Wrap and try again.

The Control Unit has detected an electronic problem in the air circuit on start-up.

Solution(s) error 02

  • Disconnect the Wrap from the Control Unit.
  • Turn the Control Unit off and on again without a Wrap connected.

The Control Unit has detected a dry pump. In order to prevent possible damage to the fluid pump, the unit will stop therapy

Solution(s) error 04

  • Make sure there is water in the ice box.
  •  Verify that the ice box filter is not clogged.
  • Make sure there are no kinks in the Wrap or Connector Hose.
  • Turn th Control Unit off and on again.
  • Disconnect and reconnect the hose from the Control Unit and the Wrap.
  • Prime the Wrap using the following steps.
  1. Select "No pressure".
  2. Attach the largest Wrap available.
  3. Lay the Wrap open (not on the body).
  4. Run the system for 1 minute.
  • Prime the System using the following steps:
  1.  Disconnect the hose from the Control Unit.
  2. Now, look at the Wrap connection location on the Control Unit. On the top valve, push the white prong in so that it is even with the metal connector.
  3. Make sure you are not fully covering the opening in the prong.
  4. Push start, and water should squirt out of the valve.
  5. Restart the System.

Indicates that the Control Unit has exceeded the target pressure

Solution(s) error 06

  • Turn the Control Unit off and back on.
  • Apply the same Wrap more tightly.
  • apply a different Wrap

The ControlUnit cant't reach its target maximum compression. This often indicates that there is a leak in the pneumatic compression circuit, either in the Connector Hose, Wrap or Control Unit. Or it may accur because the hook and loop fastener on your Wrap has worn out.

Solution(s) error 07

  • Make sure the Wrap is tightly applied
  • Try using a different Wrap and Hose to isolate the problem

The control Unit has detected that the wrap has not properly deflated

Solution(s) error 08

  • Turn the Control Unit off.
  • If there is air left in the Wrap, disconnect the Wrap from the Connector Hose and manually deflate the Wrap.
  • Reconnect the Hose to the Wrap and reapply the Wrap to the body.
  • Turn the Control Unit on and press play/pause.

The Control Unit has determined that the fluid pump may be working too hard. This could be caused by ice or debris in the fluid circuit. In order to prevent possible damage to the fluid pump, the unit will stop therapy.

Solution(s) error 09

  • Turn the Control Unit off and back on again.
  • Check to see if the filter is clogged.
  • If that does not solve the problem turn the Control Unit off for 20 minutes (to let the pump cool down) before turning it on again to try again.

Control Unit has detected an interruption in the water flow

Solution(s) error 10

  • Check all hose connections.
  • Make sure there is water in the ice box.
  • Verify that the ice box filter is bot clogged.
  • Make sure there are no kinks in the Wrap or Connector Hose.
  • Turn the Control Unit off and on again.
  • Disconnect and reconnect the hose from the Control Unit and the Wrap.

The Control Unit has detected an electronic problem in the fluid circuit on start-up.

  • Disconnect the Wrap from the Control Unit.
  • Turn the Control Unit off and on again without a Wrap connected.

Control Unit will not turn on

  • Check to make sure the AC Adapter is securely plugged in to the working electrical outlet.
  • Press the power button again. If the orange LED on the control panel is illuminated, but the Control Unit will not turn on, call the supplier. If no orange or green light is illuminated, check that the electrical outlet has power.